Specifications
| Model | AX-521 | AX-511 | AX-501 |
![]() |
![]() |
![]() |
|
| Maximum Speed | 10,000rpm | ||
| Maximum RCF | 11,740G | ||
| Maximum Capacity | 1,536ml (2ml × 96 Deepwell plate × 8) | 1,000ml (250ml × 4) | |
| Control System | Microprocessor control (feedback system) | ||
| Drive Motor | Induction motor | ||
| Drive System | Flex-Spin drive system (Direct drive with an automatic alignment function) | ||
| Data Entry | Jog Dial equipped with an <ENTER> button | ||
| Temperature Setting Range | From -9℃ to +35℃ (1℃ increment) | ||
| Speed Setting Range | 100 to 10,000rpm (100rpm increments) | ||
| RCF Setting Range | 10 to 11,740G (10G increments for the range less than 300G, 100G increments for the range over 300G) | ||
| Time Setting Range | From 10 sec to 50 sec, in 10 sec increments, from 1 min to 99 min, in 1 min increments or <F–> for free |
||
| Safety Devices | Imbalance detector, Lid interlock, Lid open / close detector, Abnormal speed detector, Over-current circuit breaker, Rotor identification system, Motor over-current detector, Abnormally high or low chamber temperature detector. |
||
| Additional Functions | Three-step acceleration characteristics selection, Three memory function, Three-step deceleration characteristics selection, FLASH (momentary spin) function, Memory function for last run parameters, BART Code selection, Power saving (ECO) mode |
||
| Refrigerant | HFC R404A (270g) | HFC R404A (260g) | HFC R404A (250g) |
| Rated Current | 120V : 12A, 220 / 230 / 240V : 6A | 120V : 10A, 220 / 230 / 240V : 5A | |
| Power Requirements | 1-Phase 50 / 60Hz AC120V 15A, 1-Phase 50 / 60Hz AC220 / 230 / 240V 8A | ||
| Power Consumption (Heat output) | 970W (834kcal / h) | 960W (826kcal / h) | 760W (654kcal / h) |
| Breaker | Rated breaking current 120V : 15A, 220 / 230 / 240V : 7.5A | ||
| Protection Against Electric Shock | Class I | ||
| Dimensions (except protruding portion) | 540W × 597D × 858H mm, (table height; 750H mm) |
510W × 567D × 858H mm, (table height; 750H mm) |
462W × 540D × 858H mm, (table height; 750H mm) |
| Net Weight | 111kg (120V) 116kg (220 / 230 / 240V) |
107kg (120V) 112kg (220 / 230 / 240V) |
99kg (120V) 104kg (220 / 230 / 240V) |
| Environmental Requirements | Ambient temperature range:10 to 35℃, Relative humidity:30 to 85%, Atmospheric pressure:700 to 1,060hPa Intended exclusively for use indoors, Overvoltage category: II, Pollution degree: 2 |
||
| Accessories Included | Operator’s manual × 1copy, BART Code table × 1pc, Clear storage case for operator’s manual × 1pc, Attaching screw for clear storage case × 1pc, Spanner × 1pc, Level gauge × 1pc, Rotor locking knob × 1pc, Drain plug × 1pc, Power cord × 1pc, Digital Warranty Information × 1copy. |
||
-
*The above specifications show the maximum value of speed, RCF (G) and sample capacity. These values may change according to the combination of rotor, bucket, adaptor and tubes.

TOP
Conical 50ml














=50ml
=15ml








![[ECO] mode](/manufactured/img/ax-img27.png)
![[ECO] key](/manufactured/img/ax-img28.png)